Section 3. Installation
Nozzle Requirements
When using a Ø 20" nozzle the height of the nozzle must not exceed 0.5 
meter. There has to be a free passage for the radar beam within a 5° angle 
from the edge of the parabolic reflector to the lower end of the nozzle. 
Nozzles with larger diameter may be higher than 0.5 meter as long as the 5° 
requirement is fulfilled.
From the flange to the product surface the distance should not be less than 
1.0 m. Highest accuracy is obtained from product levels 1.0 meter below the 
flange downwards
Figure 3-6. Nozzle requirements 
for the Rosemount 5900 with 
Parabolic antenna
Ø > 500 mm (20 inch)
Mimimum 1.0 m to 
product surface
Nozzle height 
maximum 0.5 m
5° minimum
5° minimum
Ø 441 mm
Vertical plumb line
Preliminary