Power frequency 
(50/60Hz) mag-
netic field 
IEC 61000-4-8 
3.0 A/m
3.0 A/m
Power frequency mag-
netic fields should be 
at levels characteristic 
of a typical location in 
a typical commercial or 
hospital environment.
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines IEC61000-4-11
<5% Ut (>95% dip in Ut) 
for 0.5 cycle
40% Ut (60% dip in Ut) 
for 5 cycles
70% Ut (30% dip in Ut) 
for 25 cycles
<5% Ut (>95% dip in Ut) 
for 5 sec.
<5% Ut (>95% dip in Ut) 
for 0.5 cycle
40% Ut (60% dip in Ut) 
for 5 cycles
70% Ut (30% dip in Ut) 
for 25 cycles
<5% Ut (>95% dip in Ut) 
for 5 sec.
Mains power quaility 
should be that of a typical 
commercial or hospital 
environment.  If the user 
of the EX190W image 
intensifier requires con-
tinued operation during 
power mains interrup-
tions, it is recommended 
that the EX190W image 
intensifier be powered 
from an uninterruptible 
power supply or a battery.
Note: Ut is the A.C. mains voltage prior to application of the test level.
The Model EX190W is intended for use in the electromagnetic environment specified below. 
The customer or the user of EX190W should ensure that it is used in such an environment. 
Immunity Test
IEC 60601 Test 
Level
Compliance Level
Electromagnetic environment 
- guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 
150 KHz to 80 MHz
3 V/m 
80MHz to 2.5 GHz
3 Vrms 
150 KHz to 80 MHz
3 V/m 
80MHz to 2.5 GHz
Portable and mobile RF communications 
equipment should be used no closer to any 
part of the 
EX190W, including cables, than 
the recommended separation distance 
calculated from the equation applicable to 
the frequency of the transmitter.
Recommended Separation Distance
0.15 MHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where P is the maximum oputput power 
rating of the transmitter in watts (
W
according to the transmitter manufacturer 
and d is the recommended separation 
distance in meters (
m).
Field strengths from fixed RF transmitters, 
as deter-mined by an electromagnetic 
site survey 
(a), should be less than the 
compliance level in each frequency range 
(b).Interferency may occur in the vicinity 
of equipment marked with the follwing 
symbol:
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption and
             reflection from structures, objects, and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 
radios, amateur radio, AM and FM radio broadcast and TV broadcast, connot be predicted theoretically with accuracy.  
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered.  If the measured field strength in the location in which the EX190W is used exceeds the applicable RF com-
pliance level above, the EX190W should be observed to verify normal operation. If abnormal performance is observed, 
additional measures may be necessary, such as reorienting or relocating the EX190W.
(b) Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.
Manufacturer’s declaration - electromagnetic immunity