Eizo Nanao Corporation EX240W 用户手册

下载
页码 36
30    User’s Guide
Recommended Separation Distances Between Portable and Mobile RF Communications Equi-
pement and the EX210/EX240W System
The EX210/EX240W system is intended for use in an electromagnetic environment in which radiated RF 
disturbances are controlled. The user of the EX210/EX240W system can help prevent electromagnetic 
interference by maintaining a minimum distance between portable and mobile RF communications 
equipment (transmitters) and the EX210/EX240W system as recommended below, according to the 
maximum output power of the communications equipment.
Rated maximum output 
power (W) of transmitter
Separation distance (m) according to frequency of transmitter
150kHz to 80MHz
d=1.17, p
80MHz to 800MHz
d=1.17, p
800MHz to 2.5GHz
d=2.33, p
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in me-
ters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum 
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects, and people.