viii 
Precautions about SpO
2
 
W A R N I N G  
  During SpO
2
 monitoring, always use the probe (sensor) specified by 
Fukuda Denshi. If any other probe (sensor) is used, a high temperature 
rise of the probe (sensor) may place the patient in danger of burns in 
the worst case. 
  When the SpO
2
 probe (sensor) is in a connector-off condition, the SpO
alarm will not be generated on the receiving monitor. Make sure that the 
SpO
2
 probe (sensor) is securely connected. If the SpO
waveform/numeric data is not displayed, check the patient’s condition 
and pay attention not to miss the connector-off condition. 
  When measuring the SpO
2
 of a patient with high fever or peripheral 
circulatory insufficiency, check the probe (sensor) attachment 
periodically and change the attachment site. The temperature of the 
attachment site will rise 2 to 3 C due to the sensor heat which may 
result in compression necrosis and burn injury. 
  Even a short duration of attachment may inhibit the blood flow and 
generate compression necrosis and burn injury. 
  When securing the probe (sensor) with tape, do not apply the tape too 
tight. At the same time, check the blood flow constantly so that 
congestion is not generated at the peripheral. When removing the tape, 
remove it slowly with care not to damage the patient’s skin. 
C A U T I O N  
For the following case, accurate measurement may not be possible. 
 
Patient with excessive abnormal hemoglobin (COHb, MetHb) 
 
Patient with the pigment injected to the blood 
 
Patient receiving CPR treatment 
 
Placement of SpO
2
 probe (sensor) on limb with a blood-pressure cuff, 
arterial catheter, or intravascular line 
 
When measuring at placement position with venous pulse 
 
Patient with body motion 
 
Patient with small pulse   
 
Excessive body motion (patient’s motion) 
 
Excessive light (direct sunlight, fluorescent, light therapy equipment, 
surgical light, infrared heat ramp, etc.) 
 
External colorant such as nail polish 
 
Abnormally low or high hemoglobin concentration 
 
Electrosurgery 
 
Influence of electromagnetic waves from other electronics devices 
 
High-intensity radio waves from mobile phones