Check Electrodes Screen
You can check whether the electrodes are properly attached to the patient and the ECG waveform
is acquired.
is acquired.
When 6 leads are used, I, II, Va and Vb lead waveforms are displayed.
When 3 leads are used, only II lead waveform is displayed.
Filter: off
Sweep speed:
Sweep speed:
12.5 mm/s
Waveform sensitivity: 0.5 cm/1 mV
ECG waveform
Battery level
Lead
Channel number
Detached
electrode position
electrode position
Message
When electrodes are
detached, the “CHECK
ELECTRODES”
message and detached
electrode position appear
on the screen.
detached, the “CHECK
ELECTRODES”
message and detached
electrode position appear
on the screen.
NOTE
When ECG measurement is set to OFF on the PARAMETER SETUP screen, the check
electrodes screen does not appear.
electrodes screen does not appear.