Nihon Kohden Corporation ZM-930PA 用户手册
18
Operator's Manual ZM-920PA/930PA
• All LEDs light and values are displayed on the LCD for about one second.
• The transmitter does not generate a continuous “pip” sound.
• The transmitter does not liberate excessive heat.
• The “ ” LED does not light or the “ ” mark is not displayed on the LCD.
• The transmitter does not interfere with the operation of medical instruments
• The transmitter does not generate a continuous “pip” sound.
• The transmitter does not liberate excessive heat.
• The “ ” LED does not light or the “ ” mark is not displayed on the LCD.
• The transmitter does not interfere with the operation of medical instruments
used near it.
Basic Operation
• The “signal loss” message is not displayed on the monitor when the transmitter
• The “signal loss” message is not displayed on the monitor when the transmitter
is inside the receiving range of the monitor.
• A “peep” sounds at the transmitter and “CALL” message appears at the
receiving monitor when the CALL key is pressed and the transmitter is inside
the receiving range of the monitor.
the receiving range of the monitor.
• The battery replacement message is not displayed on the monitor.
Check Items After the Power Off
• ECG electrode leads and SpO
• ECG electrode leads and SpO
2
probe are cleaned and disinfected.
• When the transmitter gets wet, liquid is wiped off and the transmitter is
thoroughly dried.
• There are enough consumables, such as disposable electrodes.
• The power is turned off.
• The batteries are removed from the transmitter when it will not be used for a
• The power is turned off.
• The batteries are removed from the transmitter when it will not be used for a
long time.
• Dead batteries are disposed of properly.