align.  The LEGI receptacle glows various colors to indicate status. See Section 6.3.10 
4.  Firmly insert the sampling line into the corresponding receptacle shown 
5.  Ensure that sampling line is secure into the monitor and not loose 
6.  Connect the appropriate airway adapter for the application as needed 
7.  Connect exhaust port to the scavenging system using part number 1846 as desired 
 
!  
CAUTION  
•  Use only accessories specifically designed and approved for use with the 
I
Radimed 
3880 system. Refer to section 9 for a complete list of available accessories. 
 
NOTE 
•  Allow the 3886 system to fully warm up (10 seconds) prior to connecting any sampling 
circuits to the patient. 
 
 
 
 
2.2.10.4. 
Internal CO2 Only Sampling Lines
 
 
The internally integrated CO2 only unit utilizes the CO2 only sample lines of Section 9.4 
This integrated CO2 system is not for use with Anesthetic agents.  The CO2 only 
sample lines have a different mating connection than those for use with the 3886 Multi-
Gas Unit and so cannot be connected to the incorrect unit. 
 
See Section 6.3.1.1 for further details of the gas sampling lines and their use.  Connect 
the CO2 only gas sampling line to the side of the 3880 Monitor Unit as shown below: