Frequency response: 
a)  Applying  a  square  wave  of  3 mV  and  100ms  to  a  recorder,  the  bias 
between the baseline after the pulse appears and the baseline before the 
pulse appears should not exceed 0.1 mV, and the slope after the pulse 
end point should be not more than 0.3 mV/s, the overshoot of the pulse 
edge should be less than 10%; 
b)For  the  sinusoidal  signal  with  a  frequency  of  0.67Hz~40Hz,  the 
response  magnitude  should  be  between  140%  and  70%  of  the 
magnitude at 5Hz (-3dB~3 dB); 
c)  The  response  amplitude  of  the  1.5mV  40ms  triangular  wave  pulse 
group for simulating a series of R wave narrowband should be between 
60%~110% of the    response amplitude at 1.5mV 200ms triangular wave 
pulse group. 
Minimum detection signal: 
When  the  operating  speed  is  set  to  25mm/s  and  the  gain  is  set  to 
10mm/mV, after a sinusoidal signal of 10Hz, 50μV(peak - Valley value) is 
applied, there should be an apparently visible deflection. 
Gain stability: 
After the device is powered on for 1min, the gain change in 3min cannot 
exceed 3% (under stable ambient conditions). 
Heart rate measurement range: 
Heart rate display range: 30bpm~200bpm, error: ±1bpm 
Measurement time 
setting 
1.    30s   
2.    180s 
Product life    5 Years 
Typical operation time 
or number of 
procedures of battery 
Battery endurance of about 8 hours 
Typical service life of 
5 years