Skin Preparation
Excessive hair, dirty skin, dry skin, or oily skin can 
impact the quality of the ECG tracing. Clean the 
patient's skin with water and soap and remove excess 
hair as needed to achieve the highest quality tracing. 
Do not use the DUO over wound areas or areas of 
broken skin.  Do not use an alcohol based skin cleaner 
as this dries out the patient's skin and increases 
resistance. Dry the skin vigorously to increase capillary 
blood flow to the tissues.
ECG gels or saline solutions can be used on the 
electrodes to improve signal quality.
FCC Intentional Radiator Certi ication
FCC ID: 2ANB3-E5
 IC:23063-E5 
This equipment contains an intentional radiator 
approved by the FCC under the FCC ID numbers 
shown above. This device complies with Part 15 of 
the FCC rules. Operation is subject to the following 
two conditions: (1) This device may not cause harmful 
interference, and (2) this device must accept any 
interference received, including interference that may 
cause undesirable operation.
NO MODIFICATION: Modifications to this device 
shall not be made without the written consent of Eko 
Devices, Inc. Unauthorized modifications may void 
the authority granted under Federal Communications 
Commission rules permitting the operation of this 
device.
EMC Compliance Europe
This equipment complies with the EMC requirements 
of the IEC 60601-1-2.
1.6
EMC Compliance
To ensure a proper ECG and PCG recording, clean the 
electrodes with an alcohol-based cleaning solution 
prior to use.
Cleaning and Disinfecting Procedure
The device and earpieces should be cleaned between 
each patient use.  All cleaning instructions pertaining 
to stethoscopes and ECG units in general apply. 
Under normal conditions it is unnecessary to separate 
the device from the earpieces for cleaning.  All external 
parts of the hardware can be cleaned with 70% 
isopropyl alcohol wipes. 
NOTE: DO NOT immerse the device in any liquid or 
subject it to any high-pressure/autoclave sterilization 
processes.
1.7
Cleaning
RSS-GEN, Sec. 7.1.3
This device complies with Industry Canada licence-exempt RSS 
standard(s). Operation is subject to the following two conditions: 
(1) this device may not cause interference, and (2) this device must 
accept any interference, including interference that may cause 
undesired operation of the device.
Le pr sent appareil est conforme aux CNR d’Industrie Canada 
applicables aux appareils radio exempts de licence. L’exploitation est 
autoris e aux deux conditions suivantes : (1) l’appareil ne doit pas 
produire de brouillage, et (2) l’utilisateur de l’appareil doit accepter 
tout brouillage radio lectrique subi, m me si le brouillage est 
susceptible d’en compromettre le fonctionnement. 
Caution: Any changes or modifications to this device not explicitly 
approved by manufacturer could void your authority to operate 
thisequipment.
Attention: Les changements ou modifications de cet appareil non 
express ment approuv par le fabricant peuvent annuler votre droit 
utiliser cet quipement.