Shenzhen Creative Industry Co. Ltd. PC-68A 用户手册
Pulse Oximeter
Directions for Use
Intended Use
This oximeter is intended to be used
with Baby Monitor, and it is an
additional or optional accessory to
measure the functional arterial oxygen
saturation (SpO
with Baby Monitor, and it is an
additional or optional accessory to
measure the functional arterial oxygen
saturation (SpO
2
) and pulse rate (PR)
for infant or baby in the way of
non-invasive and long term monitoring.
non-invasive and long term monitoring.
Contraindications
This oximeter including its sensor is
contraindicated for use on the baby with
excessive movement or for prolonged
use at the same measuring site.
contraindicated for use on the baby with
excessive movement or for prolonged
use at the same measuring site.
Instructions for Connection
1) Insert the sensor tips into the slots on the
probe holders. Place the sensor tip labeled
“LED” into the stationary probe holder on
the silicone wrap belt, and place the sensor
tip labeled “PD” into the sliding probe holder,
as illustrated in Figure 1 and Figure 2.
“LED” into the stationary probe holder on
the silicone wrap belt, and place the sensor
tip labeled “PD” into the sliding probe holder,
as illustrated in Figure 1 and Figure 2.
Figure 1 Probe tip and holder
Figure 2
Y-type SpO
2
Sensor with Rubber Wrap
NOTE:
Make sure that the silicone button
seat fully into the hole in the back of the probe
holder.
holder.
2) Bind the oximeter on the leg near ankle with
its belts and position the front panel to a
proper place for convenient operation, as
shown in Figure 4.
proper place for convenient operation, as
shown in Figure 4.
3) Wrap up the Y-type sensor onto the baby’s
sole by the rubber wrapper, as illustrated in
Figure 3. Make sure the alignment of the
light-emitting tip (LED) and the photo-
detecting tip (PD) for aiming at each other,
then thread the end of wrap belt through the
outside slots on the probe holders and tighten
the rubber belt with proper force. The final
placement is shown in Figure 4.
Figure 3. Make sure the alignment of the
light-emitting tip (LED) and the photo-
detecting tip (PD) for aiming at each other,
then thread the end of wrap belt through the
outside slots on the probe holders and tighten
the rubber belt with proper force. The final
placement is shown in Figure 4.
Figure 3 Sensor wrapping up
Figure 4
Placement of the Oximeter and the Sensor
NOTE:
The sensor connection is critical for
the signal strength and quality. Try to make
LED and PD facing each other in opposite side,
so that the light beam is as vertically
transmitted as possible and the light path is as
short as possible.