Check the Battery charge condition
Good/Bad
Good/Bad
Good/Bad
Exchange the Battery Pack with a
charged one.
charged one.
Check that the detector LED is ON.
Good/Bad
Good/Bad
Good/Bad
Set the Detector to “Exposure
Ready” as described in the
Radiography Software Manual
Ready” as described in the
Radiography Software Manual
Perform test exposure as described in
the Digital Radiography Software
Manual.
the Digital Radiography Software
Manual.
Good/Bad
Good/Bad
Good/Bad
If any error messages appear, follow
the instructions in the Digital
Radiography Software Manual. If
there is a problem, contact your
dealer, distributor, or any
PerkinElmer subsidiary.
the instructions in the Digital
Radiography Software Manual. If
there is a problem, contact your
dealer, distributor, or any
PerkinElmer subsidiary.
9.1.3
After Turning OFF the Power
Inspection
Result
Remedy
Date
/
Date
/
Date
/
G
en
er
al
Check that the XRpad™ is turned off
normally and that all LEDs are OFF
normally and that all LEDs are OFF
Good Bad
Good/Bad
Good/Bad
Check the chapter 8.7.4 for turning
off the XRpad™.
off the XRpad™.
Make sure that the XRpad™ is clean
and disinfected
and disinfected
Good Bad
Good/Bad
Good/Bad
Check the chapter 9.5 for cleaning
the XRpad™.
the XRpad™.
9.2
Monthly Inspection
Perform the following inspection at least once a month. If there is a problem, inform your
establishment safety department immediately to contact your dealer, distributor, or PerkinElmer
subsidiary.
establishment safety department immediately to contact your dealer, distributor, or PerkinElmer
subsidiary.
Inspection
Result
Remedy
Date
/
Date
/
Date
/
G
en
er
al
Execute an Image Performance Test,
and compare the test results.
and compare the test results.
Good/Bad
Good/Bad
Good/Bad
Follow the instructions in the Digital
Radiography Software Manual for
the Performance Test procedure.
Radiography Software Manual for
the Performance Test procedure.
If there are changes in the
performance, acquire new calibration
files as described in the Radiography
Software Manual.
performance, acquire new calibration
files as described in the Radiography
Software Manual.
Contact your dealer, distributor, or
any PerkinElmer subsidiaries if there
is any problem.
any PerkinElmer subsidiaries if there
is any problem.
Check the threshold of the auto trigger
mode
mode
Follow the instructions in the Digital
Radiography Software Manual for
the threshold test and if the test fails
inform your establishment safety
representative immediately to contact
your dealer, distributor, or
PerkinElmer.
Radiography Software Manual for
the threshold test and if the test fails
inform your establishment safety
representative immediately to contact
your dealer, distributor, or
PerkinElmer.
Make sure that the XRpad™ IPU is
clean from dirt or dust.
clean from dirt or dust.
Good/Bad
Good/Bad
Good/Bad
Use the instructions of chapter 0 for
cleaning.
cleaning.