Varex Imaging Deutschland AG XRPAD4336 用户手册

下载
页码 35
 
w w w . p e r k i n e l m e r . c o m  
X R p a d ™   4 3 3 6   M E D                   3 0  
 
 
 
 
 
 
 
 
 
 
 
 
U
S
E
R
 
M
A
N
U
A
L
 
12
 
 Declarations 
12.1
 
Guidance and Manufacturer’s Declaration 
Table 12  Guidance and Manufacturer’s Declaration of Electromagnetic Emissions 
Guidance and Manufacturer’s Declaration of Electromagnetic Emissions
 
The x-ray detector is intended for use in the electromagnetic environment specified below. The installer,  
x-ray system manufacturer, or user of the x-ray detector is responsible for the usage condition of the detector 
to be within such environment.
 
Emissions Test 
Compliance 
Electromagnetic Environment – Guidance 
RF-emissions CISPR 11 
Group 1 
The x-ray detector uses RF energy only for its internal 
function; therefore, its RF emissions are very low and are 
not likely to cause any interference in nearby electronic 
equipment. Should any interference (EMC) be detected 
with any other equipment, reposition the x-ray detector or 
the other equipment away from each other. 
RF-emissions CISPR 11 
Class B 
(wireless) Class 
A (wired) 
The X-Ray Detector is suitable for use in industrial and 
clinical environments in the wired mode.  
In the wireless mode the e x-ray detector is suitable for use 
in all environments within Class B.  
Should any interference (EMC) be detected with any other 
equipment, reposition the x-ray detector or the other 
equipment away from each other. 
Harmonic emissions  
IEC 61000-3-2 
Class B 
(wireless) Class 
A (wired) 
Voltage fluctuations/flicker 
emissions IEC 61000-3-3 
Complies 
 
 
Table 13  Guidance and Manufacturer’s Declaration of Electromagnetic Immunity 
Guidance and Manufacturer’s Declaration of Electromagnetic Immunity
 
The x-ray detector is intended for use in the electromagnetic environment specified below. The installer,  
x-ray system manufacturer, or user of the x-ray detector is responsible for the usage condition of the detector 
to be within such environment.
 
Immunity Test
 
IEC 60601 Test
 
Compliance
 
Electromagnetic Environment – 
Guidance
 
Electrostatic Discharge 
(ESD)  
IEC 61000-4-2
 
Contact: 6 kV 
Air: 8 kV
 
Contact: 6 kV 
Air: 8 kV
 
Floors should be made of wood, 
concrete, or ceramic tile. If floors are 
covered with synthetic material, the 
relative humidity should be at least 30%.
 
Electrical fast transients 
(Burst) IEC 61000-4-4 
0.5 kV (AC) 
1kV (DC) 
0.5 kV (AC) 
1kV (DC) 
Mains power quality should be that of a 
typical commercial and/or hospital 
environment. 
Transients-Surges 
IEC 61000-4-5 
1 kV / 2 kV 
1 kV / 2 kV 
Mains power quality should be that of a 
typical commercial and/or hospital 
environment. 
Power frequency magnetic 
field 
IEC 61000-4-8 
3 A/m 
3 A/m 
Power frequency magnetic fields should 
be at levels characteristic of a typical 
location in a typical commercial and/or 
hospital environment. 
Voltage dips and short 
interruptions 
IEC 61000-4-11 
-95% / 10 ms 
-60% / 100 ms 
-30% / 500 ms 
>-95% / 5000 
ms 
-95% / 10 ms 
-60% / 100 ms 
-30% / 500ms 
>-95% / 5000 
ms 
Mains power quality should be that of a 
typical commercial or hospital 
environment. 
If the user of the x-ray detector requires 
continued operation during power mains 
interruptions, we recommend that the x-
ray detector be powered from an 
uninterruptible power supply or battery.