MEDTRUM TECHNOLOGIES Inc. JN001 用户手册
37
14. After you fill the Reservoir Patch, the reservoir volume icon will change from empty to full, and then press OK.
Note: The reservoir volume icon will change to full if you have filled the Reservoir Patch with at least 70 units of
insulin. If you have filled the Reservoir Patch with more than 70 units and the reservoir volume icon is empty, call
Customer Care.
15. The system automatically primes the Reservoir Patch.
Note: After filling the Reservoir Patch with insulin, you should complete the Reservoir Patch change process within 3
hours. As a reminder that the Reservoir Patch has been filled, it will beep and/or vibrate every 10 minutes to indicate
that time is passing. If you do not set up the Reservoir Patch within 3 hours, you must deactivate and discard it.
Note: Once a Reservoir Patch (assembled with Pump Base) is activated and communicates with the PDM, it can
only receive commands from that PDM, not from any other.
Select and prepare the infusion site
Warning: The place on your body where you attach the Reservoir Patch is important for the success of your therapy.
Choose a location that remains flat when you are sitting down, standing up, or lying down. Discuss the best location
for you with your doctor or healthcare professional.
Shown here are the best body areas (shaded) for infusion sites. The abdominal area is the most common site
because absorption is very consistent. Other potential sites may be used, that typically have a layer of fatty tissues,
such as the hip, back of upper arm, upper thigh, or lower back.