Version 02.01 • Date 19JAN2017
A PEF measurement performed with the AM3 is valid, if:
•  Breathing volume > 0.47 L or < 10 L
•  Breathing flow > 50 L/min
•  FVC > FEV1
If all efforts are invalid, no measurement data will be stored on the device.
4.3  Validity of PEF Measurements
To store a result in the AM3, at least one adequate maneuver needs to be 
performed.
OK
ESC
XX
Otherwise a message will be displayed and you 
will be asked to repeat the measurement.
4.4  Performing Scheduled Sessions
Scheduled sessions, including questionnaire and PEF measurements, can only be 
performed and stored using programmed devices.  
For completing your questionnaire, you can scroll through and select the appropriate 
answer by using 
 and 
 buttons and pressing 
OK
.
At the end of the questionnaire you will be asked, if you want to change an answer; 
if you select “Yes“, the questionnaire will be displayed again. The answers previously 
entered will appear as default, and can be modified as described above.