eResearchTechnology GmbH AM3G02 用户手册
Page 17/40
Version 02.01 • Date 19JAN2017
A PEF measurement performed with the AM3 is valid, if:
• Breathing volume > 0.47 L or < 10 L
• Breathing flow > 50 L/min
• FVC > FEV1
• Breathing volume > 0.47 L or < 10 L
• Breathing flow > 50 L/min
• FVC > FEV1
If all efforts are invalid, no measurement data will be stored on the device.
4.3 Validity of PEF Measurements
To store a result in the AM3, at least one adequate maneuver needs to be
performed.
performed.
OK
ESC
XX
Otherwise a message will be displayed and you
will be asked to repeat the measurement.
will be asked to repeat the measurement.
4.4 Performing Scheduled Sessions
Scheduled sessions, including questionnaire and PEF measurements, can only be
performed and stored using programmed devices.
For completing your questionnaire, you can scroll through and select the appropriate
answer by using
performed and stored using programmed devices.
For completing your questionnaire, you can scroll through and select the appropriate
answer by using
and
buttons and pressing
OK
.
At the end of the questionnaire you will be asked, if you want to change an answer;
if you select “Yes“, the questionnaire will be displayed again. The answers previously
entered will appear as default, and can be modified as described above.
if you select “Yes“, the questionnaire will be displayed again. The answers previously
entered will appear as default, and can be modified as described above.