7.  Error Checklist
Error Description
Reason
Action
No response during 
power ON
A.
AM3 battery is empty
Charge the AM3
B.
Buttons 
OK
 and 
ESC
 are not pressed 
correctly
•  Press and hold the 
OK
 button.
•  While holding the 
OK
 button,  
 
press and hold the 
ESC
 button.
•  Hold both buttons 
for approximately  
2 seconds.
Result of 
measurements is 
questionable
A.
Flow Sensor is not 
inserted correctly
Insert Flow Sensor 
correctly
B.
Flow Sensor is dirty
Clean Flow Sensor 
according to cleaning 
instructions
C.
Flow Sensor is faulty
Replace Flow Sensor
If the proposed actions do not lead to perfect recovery of the AM3´s 
normal functionality, please contact your study doctor.