eResearchTechnology GmbH AM3G02 用户手册

下载
页码 40
Page 36/40
Version 02.01 • Date 19JAN2017
10.  Technical Data
Principle:
Determination of respiratory flow and volume via 
exchangeable infrared rotary flow sensor.
Range:
Flow
Volume
0 - 840 L/min
0.5 - 8 L
Accuracy:
Flow
Volume
± 5 % or ± 20 L/min
± 3 % or ± 0.05 L
Storage capacity: 
1200 measurements, 400 sets of questionnaires  
(max. 20 questions each)
Power supply:
Built-in rechargeable lithium-ion battery 3.7 V, 1700 mAh 
Battery will last under standard operating conditions for 
about 40 days.   
Full charging:  2 h 
Cycle life:  
80% of rated capacity after 300 cycles 
 
60% of rated capacity after 500 cycles
Dimensions:
Length x width x height
Weight
112 x 82 x 37 mm
150 g (battery included)
Ambient conditions:
Temperature
Relative humidity
Barometric pressure
+10 °C to +40 °C
15 % to 95 %, not condensing
700 to 1060 hPa
Transport and storage 
conditions:
Temperature
Relative humidity
Barometric pressure
-20 °C to +50 °C
15 % to 95 %, not condensing
600 to 1200 hPa
Moisture protection:
IP 22
Medical classification:
Active Medical Device Class IIa