The sensor is intended for single patient use, only. If the AM3 is passed 
on to another patient, the AM surfaces will have to be cleaned and 
disinfected with a disinfectant. See 10.4 for cleaning of housing. 
See below for informa tion on disposal of a used sensor.
10.  Cleaning
10.1  Cleaning of Sensor
To clean the rotary flow sensor, release and remove sensor, rinse it with mild 
cleansing agent, e.g. Instruton E by ANTISEPTICA Dr. Hans-Joachim Molitor GmbH 
according to instructions of manufacturer, and then rinse with clean distilled water. 
Shake off any remaining water on the sensor. Air-dry and reinsert the sensor. Any 
minor discoloration in the sensor does not affect the performance of the AM3.
The Cleaning of the sensor should be performed every week or more frequently if 
often heavily polluted. Furthermore, it is recommended to exchange the rotary flow 
sensory with a new one after one year of use or 2000 measurements. If the sensor 
is heavily polluted frequently, the exchange with a new sensor should be performed 
every six months or as required. If there are any doubts that the measurement values 
are correct, the rotary flow sensor has to be exchanged immediately.
DO NOT use alcohol or any type of household cleaner!
10.2  Checking the Sensor
If your AM3 does not measure accurately, exchange the sensor or contact the 
responsible Monitor or ERT Customer Care Helpdesk.