eResearchTechnology GmbH AM3G03 用户手册

下载
页码 40
Page 37/40
Version 01.00 • Date 25JAN2018
12.  Technical Data
Principle:
Determination of respiratory flow and volume via exchangeable 
infrared rotary flow sensor.
Range:
Flow
Volume
0 - 840 l/min
0.5 - 8 l
Accuracy:
Flow
Volume
± 5 % or ± 20 l/min
± 3 % or ± 0.05 l
Storage capacity: 
1200 measurements, 400 sets of questionnaires  
(max. 20 questions each)
Power supply:
Built-in rechargeable lithium-ion battery 3.7 V, 1700 mAh 
Battery will last under standard operating conditions for about 40 days.   
Full charging: 
2 h 
Cycle life:  
80% of rated capacity after 300 cycles 
 
60% of rated capacity after 500 cycles
Dimensions:
Length x width x height
Weight
112 x 82 x 37 mm
150 g (battery included)
Ambient conditions:
Temperature
Relative humidity
Barometric pressure
+10 °C to +40 °C
15 % to 95 %, not condensing
700 to 1060 hPa
Transport and storage 
conditions:
Temperature
Relative humidity
Barometric pressure
-20 °C to +50 °C
15 % to 95 %, not condensing
600 to 1200 hPa
Moisture protection:
IP 22
Medical classification:
Active Medical Device Class IIa
Applied part:
Type BF (whole device)
Protection class:
Battery Device
Mode of operation:
Continuous operation
Max. resistance:
0.002 kPa/l/min at 720 l/min
Interface:
USB 2.0 
Bluetooth 4.2 Smart Ready
Medical Power supply 
(battery charging):
WR9QA1200MUNMRVG2773                                      
Model  GTM41134-0605
Input     
100-240 Vac,  50/60 Hz, 0.3 A
Output   
5 V, 1.2 A
Cable length    1500 mm
The expected operational lifetime of the AM3 is 5 years. 
AM corresponds to the recommendations of ATS/ERS.