TROUBLESHOOTING
The following table lists the recommended actions for potential issues with the CMCT
System. Please call ZOLL at +972-9-9603900 if you need assistance with any of these
instructions
.
Signs
Possible Causes
Actions
Sensor light indicator
is not on while sitting
in the Charger for
charging.
The Sensor does not sit
properly in the Charger.
Remove the Sensor from the Charger and
reinsert back to the Charger.
The Charger is not connected
to the power outlet.
Plug in the Charger to the power outlet using
the power cable provided in the CMCT
packaging. When properly connected, the
light indicator on the Charger should be on
and is green.
Sensor light turns
blinking amber after
being applied on the
patient’s body.
Sensor Error.
Contact ZOLL customer service
at
1.888.592.3798.
Charger light indicator
is not on.
The Charger is not connected
to the power outlet.
Plug in the Charger to the power outlet using
the power cable provided in the CMCT
packaging. When properly connected, the
light indicator on the Charger should be on
and is green.
The Gateway screen
does not indicate the
Gateway is being
charged while sitting
in the Charger for
charging.
The Gateway does not sit
properly in the Charger.
Remove the Gateway from the Charger and
reinsert back to the Charger.
7
CLEANING AND RETURN OF SYSTEM
Users are not required to clean the CMCT System. When the monitoring period is over, use
the box provided in the original packaging to return the CMCT System (Sensor, Charger,
Gateway, power cable, and any unused Patches) to ZOLL.