5.6. REMOVING THE DEVICE CONNECTIONS FROM THE PATIENT 
The Cartridge should be removed from the Medical Patch as follows: 
1.
 
Disconnect the two µCor ESS leads from the Medical Patch. 
2.
 
Tilt and remove the Cartridge from the Medical Patch while pressing the upper 
snap in clip. 
3.
 
Remove the Medical Patch from the 
patient. 
4.
 
Place the Cartridge in the charger and 
wait for a steady amber LED light. (If no 
light appears after 1-2 seconds check 
troubleshooting table for required actions)  
5.
 
If patient is still active, the amber light will change to green after 1-2 hours, 
indicating the device is fully charged. 
6.
 
If the patient has been discharged, the steady amber light should change to a 
flashing green light within around two hours, indicating the device is charged 
and ready for a new patient.  
 Note:  A steady (non-flashing) green light means the µCor is still 
referenced to the previous patient at the server.  
The Cartridge is ready for a new patient only when there is a flashing 
green LED light- DO NOT APPLY DEVICE TO A NEW PATIENT UNTIL THIS 
INDICATION OCCURS. 
The Cartridge should be cleaned and disinfected before using it for a new patient. Please see 
cleaning and disinfecting instructions on page 22. 
 
5.7 CLEANING THE CARTRIDGE  
After disconnecting the Cartridge from the patient: 
 
Clean and disinfect the Cartridge, as shown on page 22. 
 
Place the Cartridge in the charger and make sure the LED light is on. 
When the green LED light is flashing, the system is ready for a new patient. 
(previous patient is released and the battery is fully charged) 
 
 
Caution: If the green LED light is not flashing, the last active 
patient was not yet released from the server. 
 
 
 
Steady amber light 
Flashing green light