DTE4027 RFID Reader Module 
User Manual 
 
4  INSTALLATION PROCEDURE 
 
 
4.1  Installation Requirements 
 
In  order  to  reduce  unwanted  emission  in  the  spurious  domain  under  the  maximum  regulatory  limits,  two 
ferrites must be added on the cable connecting the RFID board to the instrument at a distance of 70 mm 
maximum from the RFID board connector. The ferrites are mounted as described: 
 
 
One ferrite on the power supply wires 
 
One ferrite on the data and switch wires 
 
Wires must make a loop one time through the ferrite 
 
Ferrites must be Würth Elektronik P/N 7427114  
 
4.2  Installation 
 
Install  the  board  with  screws  using  the  four  holes.  The  PCB  should  be  covered  by  an  RF  permissive 
material cover (e.g. plastic). 
 
Power the board with a regulated 5 VDC power supply. 
 
4.3  Command set 
 
The  command  set  is  described  in  the  MFPROT_LP.pdf  document  that  can  be  downloaded  from 
http://www.ibtechnology.co.uk/pdf/MFprot_LP.PDF
. Please refer to this document to control the board. 
 
 
5  TROUBLESHOOTING 
 
 
5.1  Error Messages 
 
 
Error Message 
Probable Cause(s) 
LED doesn’t light up 
Check the presence of 5VDC between pin 1 and 2 of the 
connector 
The board doesn’t respond to UART 
Check the host baud rate 
RFID Tag cannot be red 
Check the tag type and distance to the reader