3. The rule of Multi-Share control is based on IP address. The second terminal with different IP address is
not allowed to operate panel after the first one connected. If there is no command transmission between panel
and Workstation over 5 minutes, panel releases access authority.
4.3 Panel Configuration
Click “Config”
Set parameters
Note: 1. If panel works in Isync Plus mode, it is not allowed to change any parameters and write into ROM
or RAM. User is required to switch to software mode, change parameters and then switch to Isync Plus
mode. On the other hand, we do not recommend user to switch working mode too often.
4.4 Correction and Calibration Template Generation
Iray recommends performing correction and calibration after installation or any major change on the system
settings and hardware configuration. On the other hand, it is also recommended to do the correction and
calibration in each 6 months.
4.4.1 Pre‐offset Template Generation
If panel is configured to do Pre-offset correction, Pre-offset Template is necessary. See below