ResMed A10STA3G 用户手册
14
Problem/possible cause
Solution
Air pressure in my mask seems too low (it feels like I am not getting enough air)
Ramp may be in progress
.
Wait for air pressure to build up or turn Ramp Time off.
Ramp Down may be in progress
.
Press Start/Stop to stop therapy then press Start/Stop to
restart and continue therapy.
Non-vented mask is used.
Only use a vented mask.
Mask vents might be blocked.
Check if you have sufficient venting. Unblock mask vents if
necessary.
Expiratory pressure (EPAP) may be set too low.
Talk to your care provider about your settings.
My screen is black
Backlight on the screen may have turned off. It turns off
automatically after a short period of time.
Press Home or the dial to turn it back on.
Power may not be connected.
Connect the power supply and make sure the plug is fully
inserted.
Note: The retention clip should be in the open position when
inserting the plug. For instructions refer to the Setup section.
I have stopped therapy, but the device is still blowing air
Device is cooling down.
Device blows a small amount of air in order to avoid
condensation in the air tubing. It will stop automatically after
condensation in the air tubing. It will stop automatically after
20 minutes.
My water tub is leaking
Water tub may not be assembled correctly.
Check for damage and reassemble the water tub correctly.
Water tub may be damaged or cracked.
Contact your care provider for a replacement.
My therapy data has not been sent to my care provider
Power may not be connected.
Connect the power supply and make sure the plug is fully
inserted.
Note: The retention clip should be in the open position when
inserting the plug. For instructions refer to the Setup section.
Wireless coverage may be poor.
Make sure that the device is placed where there is coverage
(ie, on your bedside table, not in a drawer or on the floor).
The Wireless signal strength icon indicates good
(ie, on your bedside table, not in a drawer or on the floor).
The Wireless signal strength icon indicates good
coverage when all bars are displayed, and poor coverage
when fewer bars are displayed.
The No wireless connection icon
is displayed on the
top right of the screen. no wireless network available.
Make sure that the device is placed where there is coverage
(ie, on your bedside table, not in a drawer or on the floor).
(ie, on your bedside table, not in a drawer or on the floor).
If instructed to do so, send the SD card to your care provider.
The SD card also contains your therapy data.
Device may be in Airplane Mode.
Turn off Airplane Mode, see Traveling by plane.