Problem/possible cause 
Solution 
Air pressure in my mask seems too low (it feels like I am not getting enough air) 
Ramp may be in progress 
Wait for air pressure to build up or turn Ramp Time off. 
Ramp Down may be in progress 
Press Start/Stop to stop therapy then press Start/Stop to 
restart and continue therapy. 
Non-vented mask is used. 
Only use a vented mask. 
Mask vents might be blocked. 
Check if you have sufficient venting. Unblock mask vents if 
necessary. 
Expiratory pressure (EPAP) may be set too low. 
Talk to your care provider about your settings. 
My screen is black 
Backlight on the screen may have turned off. It turns off 
automatically after a short period of time. 
Press Home or the dial to turn it back on. 
Power may not be connected. 
Connect the power supply and make sure the plug is fully 
inserted. 
Note: The retention clip should be in the open position when 
inserting the plug. For instructions refer to the Setup section. 
I have stopped therapy, but the device is still blowing air 
Device is cooling down. 
Device blows a small amount of air in order to avoid 
condensation in the air tubing. It will stop automatically after 
20 minutes. 
My water tub is leaking 
Water tub may not be assembled correctly. 
Check for damage and reassemble the water tub correctly. 
Water tub may be damaged or cracked. 
Contact your care provider for a replacement. 
My therapy data has not been sent to my care provider 
Power may not be connected. 
Connect the power supply and make sure the plug is fully 
inserted. 
Note: The retention clip should be in the open position when 
inserting the plug. For instructions refer to the Setup section. 
Wireless coverage may be poor. 
Make sure that the device is placed where there is coverage 
(ie, on your bedside table, not in a drawer or on the floor). 
The Wireless signal strength icon   indicates good 
coverage when all bars are displayed, and poor coverage 
when fewer bars are displayed. 
The No wireless connection icon 
 is displayed on the 
top right of the screen. no wireless network available. 
Make sure that the device is placed where there is coverage 
(ie, on your bedside table, not in a drawer or on the floor). 
If instructed to do so, send the SD card to your care provider. 
The SD card also contains your therapy data. 
Device may be in Airplane Mode. 
Turn off Airplane Mode, see Traveling by plane.