Connecting HF instruments
1.  Refer to the instructions for use of the HF instrument.
2.  Inspect the HF instrument, its connection cable and plug for scratches, cracks, and other 
damages.
Do not use the HF instrument if any kind of damage is observed.
3.  Properly insert the plug of the HF instrument into the corresponding output socket on the 
front panel of the electrosurgical generator.