Clarity - INT9 A/D converter 
 Installation 
Starting cable
Signal cable
START (REMOTE, …)
Red
Shielding
SYMMETRICAL CONNECTION
+    (+1V, Hi, …)
-     (-1V, Lo, …)
COM (0, GND)
Red
Shielding
White
Digital output (Relay)
START (REMOTE, …)
1V  (100mV, Hi, +, …)
0    (COM, AGND, …)
NONSYMMETRICAL CONNECTION
Red
Shielding
Red
Shielding
White
Starting cable
Signal cable
Digital output (Relay)
 
Fig. 2. 
Connecting Clarity to chromatograph 
2.3.1 
Connecting signal cables: 
Signal inputs of the A/D converters are 
symmetrical: + (red), - (white) and analogue 
ground (copper braiding). 
Caution! 
Shielding must be connected. It serves not only as 
the shielding but also as the analogue earth 
against which the measurement takes place. In 
the case of asymmetrical output of a detector (only 
two connectors) the shielding must be connected 
to the white lead! No lead of the signal cable may 
remain unconnected. 
Try to connect to the detector output of the 
chromatograph with the largest possible signal 
level, usually indicated as INTEGRATOR 
(signal approx. 1V). The level of the signal on 
the output marked RECORDER is usually only 
approx. 10mV. 
For easier alterations of the wiring, we supply a 
SV8  Terminal board with screw contacts for 
INT7,  INT9,  U-PAD, U-PAD2 and  Net-PAD 
A/D converters. 
Note: 
This SV8 Terminal Board is not suitable for 
applications with small signals or with high 
electromagnetic interference. This is because the SV8 
Terminal Board uses leads and screw contacts that 
are not shielded.