Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S 用户手册

产品代码
MSO1074Z-S
下载
页码 172
RIGOL 
2-110 
  DS1000Z Programming Guide 
:TRIGger:WINDows:POSition 
Syntax 
:TRIGger:WINDows:POSition <pos> 
:TRIGger:WINDows:POSition? 
Description 
Set or query the trigger position in windows trigger. 
Parameter 
Name 
Type 
Range 
Default 
<type> 
Discrete 
{EXIT|ENTER|TIMe} 
ENTER 
 
Explanation 
  EXIT: trigger when the input signal exits the specified trigger level range.
 
  ENTER: trigger when the trigger signal enters the specified trigger level range.
 
  TIMe: used to specify the hold time of the input signal after entering the specified 
trigger level range. The instrument triggers when the accumulated hold time equals 
the windows time.
 
Return 
Format 
The query returns EXIT, ENTER or TIM.   
Example 
:TRIGger:WINDows:POSition ENTER 
 
 
:TRIGger:WINDows:TIMe 
Syntax 
:TRIGger:WINDows:TIMe <NR3> 
:TRIGger:WINDows:TIMe? 
Description 
Set or query the hold time in windows trigger. 
Parameter 
Name 
Type 
Range 
Default 
<NR3> 
Real 
8ns to 10s 
1μs 
 
Return 
Format 
The query returns the hold time in scientific notation.   
Example 
:TRIGger:WINDows:TIMe 0.002   /*Set the hold time to 2ms*/ 
:TRIGger:WINDows:TIMe?   /*The query returns 2.000000e-03*/ 
Related 
Command