Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S 用户手册

产品代码
MSO1074Z-S
下载
页码 172
RIGOL 
DS1000Z Programming Guide 
2-117
 
:TRIGger:SHOLd:STIMe 
Syntax 
:TRIGger:SHOLd:STIMe <NR3> 
:TRIGger:SHOLd:STIMe? 
Description 
Set or query the setup time in setup/hold trigger. The default unit is s. 
 
Parameter 
Name 
Type 
Range 
Default 
<NR3> 
Real 
8ns to 1s 
1μs 
 
Explanation 
This command is only available when the setup type (refer to 
) is 
SETup or SETHOLd. 
 
Return 
Format 
The query returns the setup time in scientific notation.   
Example 
:TRIGger:SHOLd:STIMe 0.002   /*Set the setup time to 2ms*/ 
:TRIGger:SHOLd:STIMe?   /*The query returns 2.000000e-03*/ 
Related 
Command 
 
 
:TRIGger:SHOLd:HTIMe 
Syntax 
:TRIGger:SHOLd:HTIMe <NR3> 
:TRIGger:SHOLd:HTIMe? 
Description 
Set or query the hold time in setup/hold trigger. The default unit is s. 
 
Parameter 
Name 
Type 
Range 
Default 
<NR3> 
Real 
8ns to 1s 
1μs 
 
Explanation 
This command is only available when the setup type (refer to 
) is 
HOLd or SETHOLd. 
 
Return 
Format 
The query returns the hold time in scientific notation.   
Example 
:TRIGger:SHOLd:HTIMe 0.002   /*Set the hold time to 2ms*/ 
:TRIGger:SHOLd:HTIMe?   /*The query returns 2.000000e-03*/ 
Related 
Command