  DS1000Z Programming Guide 
IEEE488.2 Common Commands 
 
IEEE 488.2 standard defines some common commands used for querying the basic information of the 
instrument or executing basic operations. These commands usually start with "*" and the keyword of the 
command is usually 3-character long.   
 
Command List: 
  *CLS 
  *ESE 
  *ESR? 
  *IDN? 
  *OPC 
  *RST 
  *SRE 
  *STB? 
  *TST? 
  *WAI 
 
 
*CLS 
Syntax 
*CLS 
Description 
Clear all the event registers in the register set and clear the error queue. 
 
 
*ESE 
Syntax 
*ESE <value> 
*ESE? 
Description 
Set or query the enable register for the standard event register set. 
Parameter 
Name 
Type 
Range 
Default 
<value> 
Integer 
0~255 
 
Explanation 
The bit 1 and bit 6 of the standard event status register are not used and are always 
treated as 0, therefore, the range of <value> are the decimal numbers corresponding to 
the binary numbers ranging from 00000000 (0 in decimal) to 11111111 (255 in decimal) 
and of which the bit 1 and bit 6 are 0. 
Return 
Format 
The query returns an integer which equals the sum of the weights of all the bits that have 
already been set in the register. 
Example 
*ESE 16    /* Enable the bit 4 (16 in decimal) of the standard event status register */ 
*ESE?     /*The query returns 16*/