  DS2000A User’s Guide 
6-6 
to set the scale and position of the horizontal axis respectively. 
The unit of the vertical axis can be dB or Vrms which use logarithmic mode and 
linear mode to display vertical amplitude respectively. If you need to display the 
FFT frequency spectrum in a relatively larger dynamic range, dB is 
recommended. Press Scale to select the desired unit and the default is Vrms. 
Press 
  and 
  and use 
  to set the vertical position and amplitude of 
the FFT frequency spectrum respectively. 
 
Tips 
 
Signals with DC components or deviation would cause error or deviation of the FFT 
waveform components. To reduce the DC components, set the 
to “AC”. 
To reduce the random noise and aliasing frequency components of repetitive or 
single pulse, set the 
 of the oscilloscope to “Average”.