DS2000A User’s Guide
 
 
3-7  
 
Roll Mode 
 
In this mode, the waveform scrolls from the right to the left to update the display and 
the waveform horizontal position and trigger control are not available. The range of 
horizontal scale adjustment is from 200.0 ms to 1000 s.   
 
Note: When Roll mode is enabled, 
, 
 and 
 are not available.   
 
Slow Sweep 
 
Another mode similar to Roll mode. When the horizontal time base is set to 200 ms 
or slower, the instrument enters “slow sweep” mode in which the instrument first 
acquires the data at the left of the trigger point and then waits for trigger event. 
After the trigger occurs, the instrument continues to finish the waveform at the right 
of the trigger point. When slow sweep mode is used to observe low frequency 
signal, it is recommended that the 
 is set to “DC”.