DS2000A User’s Guide
 
 
5-11  
 
Pulse Width Setting:   
In this oscilloscope, positive pulse width is defined as the time difference between 
the two crossing points of the trigger level and positive pulse as shown in the figure 
below. 
 
 
 
 
 
 
 
  When the Pulse Condition is set to 
  or 
, press 
Setting and use 
  to input the desired value. The range available is from 2 ns 
to 4 s. 
  When the Pulse Condition is set to 
  or 
, press Upper Limit 
and Lower Limit and use 
  to input the desired values respectively. The 
range of the upper limit is from 10 ns to 4 s. The range of the lower limit is from 
2 ns to 3.99 s. Note that the lower limit of the pulse width must be lower than 
the upper limit. 
 
Trigger Mode: 
Press Sweep to select the 
 (page 
) under this trigger type as auto, 
normal or single. The corresponding status light of the current trigger mode turns on.   
 
Trigger Setting: 
Press Setting to set the trigger parameters (trigger holdoff and noise rejection) 
under this trigger type.   
 
Trigger Level: 
Use TRIGGER 
  LEVEL to modify the level. For details, please refer to the 
description of 
 on page 
.   
 
 
 
 
 
 
Trigger level 
 
Positive 
pulse width