DS2000A User’s Guide
 
 
10-9  
 
the progress bar is displayed and the parameters can not be modified. After the 
analysis finishes, the analysis results of “Error Frames”, “Current Error” and 
“CurFrame Diff” are displayed; at the same time, the first error frame is located 
as shown in the figure below. At this point, the next error frame, the previous 
error frame as well as every frame in this waveform analysis can be located 
using the corresponding softkeys.   
 
 
 
Error Frames:   
The total number of error frames discovered in this analysis. The number of 
error frames is related to the error frame threshold currently set. 
 
Current Error:   
The order number of the error frame lately located in all the error frames when 
locating every data frame in this analysis.   
 
CurFrame Diff:   
During the analysis based on trace, the oscilloscope compares each frame with 
the templet to compute the difference value and normalizes each value using 
the maximum one; then, compares the normalized value of each frame with the 
threshold selected to judge whether the frame is an error frame. “CurFrame Diff” 
(relative difference) is the normalized value of the difference between the 
current frame and templet. 
 
During the analysis based on pass/fail mask, the oscilloscope compares each 
frame with the mask to compute the difference value and recognizes the frame 
whose difference value is equal to or larger than the threshold selected as an 
error frame and the corresponding “CurFrame Diff” is 100%; otherwise, the 
frame is juged as correct and the “CurFrame Diff” is 0%. Note that there are 
only two “CurFrame Diff” values (100% and 0%) under the analysis based on 
pass/fail mask.