5-10 
knob, the trigger level line and the trigger mark disappear in about 2 s. 
 
 
Pulse Trigger 
 
Trigger on the positive or negative pulse with a specified width.   
 
Trigger Type: 
Press Type to select “Pulse”. At this point, the trigger setting information as shown in 
the figure below is displayed at the upper right corner of the screen.   
 
 
Source Selection: 
Press Source to select CH1 or CH2 as the 
. The current trigger 
source is displayed at the upper right corner of the screen. 
Note: Select channel with signal input as trigger source to obtain stable trigger.   
 
Pulse Condition:   
Press When to select the desired pulse condition. 
 
: trigger when the positive pulse width of the input signal is greater than 
the specified pulse width. 
 
: trigger when the positive pulse width of the input signal is lower than the 
specified pulse width. 
 
: trigger when the positive pulse width of the input signal is greater than 
the specified lower limit of pulse width and lower than the upper limit of pulse 
width. 
 
: trigger when the negative pulse width of the input signal is greater than 
the specified pulse width. 
 
: trigger when the negative pulse width of the input signal is lower than 
the specified pulse width. 
 
: trigger when the negative pulse width of the input signal is greater than 
the specified lower limit of pulse width and lower than the upper limit of pulse 
width.