Trigger Coupling   
 
Trigger coupling decides which kind of components will be transmitted to the trigger 
circuit. Please distinguish it with 
.   
  DC: allow DC and AC components into the trigger path.  
  AC: block all the DC components and attenuate signals lower than 8 Hz. 
  LF Reject: block the DC components and reject the low frequency components 
(lower than 75 kHz). 
  HF Reject: reject the high frequency components (higher than 75 kHz). 
 
Press MENU  Setting  Coupling in the trigger control area (TRIGGER) at the 
front panel to select the desired coupling type (the default is DC). Note that trigger 