Rohde Schwarz Rohde & Schwarz HMO1002 2-channel oscilloscope, Digital Storage oscilloscope, 21-1002-RS00 数据表

产品代码
21-1002-RS00
下载
页码 60
34
Signal  Generation
10  Signal  
Generation
10 .1  Function Generator
The R&S®HMO1002 includes an integrated function ge-
nerator which can generate input signals while testing 
circuits, for instance. To activate the function generator, 
select the UTIL menu in the VERTICAL control panel, and 
then press the soft menu key FUNCTION GEN. The fol-
lowing waveforms can be generated and made available at 
the AUX OUT connector:
 DC
 
SINE: Frequency range from 0.1 Hz to 50 kHz
 SQUARE WAVE: Frequency range from 0.1 Hz to 50 kHz
 PULSE: Frequency range from 0.1 Hz to 10 kHz
 TRIANGLE: Frequency range from 0.1 Hz to 10 kHz
 RAMP: Frequency range from 0.1 Hz to 10 kHz
Use the soft menu key FREQUENCY to select the signal 
frequency, and use the soft menu key AMPLITUDE to se-
lect the signal amplitude via universal knob or numerically 
via KEYPAD key in the CURSOR/MENU control panel. 
You also have the option to select a DC offset (soft menu 
key OFFSET). An information box shows a preview of the 
respective signal form with the selected parameters.
Use the soft menu key OFF to deactivate the function ge-
nerator and to close the menu.
10 .2  Pattern Generator
The pattern generator includes the output of parallel pat-
terns on the four pins S0 to S3 on the front panel of the 
instrument. It is based on a 2048 bit wide memory (Sam-
ples) which can be issued cyclically / individually. All sub 
functions of the PATTERN GENERATOR are based on this 
memory. It is not possible to issue multiple functions simul-
taneously. An information box shows which pins to use.
The pattern generator menu (pattern generator) includes 
various options. Use the UTIL menu in the VERTICAL con-
trol panel and press the soft menu key PATTERN GEN. to 
open it.
Function
Setting options
Square wave
Frequency / Period, Polarity, Duty Cycle
Counter
Frequency, counting direction
Arbitrary
Timing, pattern input
Manual 
Manual switching of the 4 individual 
connectors
UART
Polarity, setting the bit rate
SPI / I
2
C / CAN / LIN
Setting the bit rate
10 .2 .1  Square Wave
Use the soft menu SQUARE WAVE to perform a manual 
probe compensation without compensation wizard. The 
Tab. 10.1: Settings options for the pattern-generator
square wave function uses a 100-sample wide pattern (100 
conditions). Additionally, you can change the polarity and 
the duty cycle (1% to 99%).
10 .2 .2 Counter
A 4 bit wide counter pattern is is-
sued. The user can determine the 
counting direction (soft menu key 
DIRECTION) and the frequency (soft 
menu key FREQUENCY). The user 
frequency always refers to the swit-
ching of the pattern condition. This 
results in square waveforms for individual pins, as descri-
bed in tab. 10.2.
10 .2 .3 Arbitrary
The user can define a 4 bit wide and a 2048 sample deep
pattern via soft menu ARBITRARY. The created patterns 
can be stored or recalled. If the arbitrary function is activa-
ted, the predefined pattern, which may have been gene-
rated automatically, will be retained. This means that a SPI 
pattern can be analyzed and adapted as necessary. 
.
The following settings options are available:
 PATTERN SETUP:  
Use the soft menu key PATTERN LENGTH to determine 
the pattern depth. Use the INDEX and the universal knob 
or the KEYPAD key in the CURSOR/MENU control panel 
to select individual samples. The selected sample will be 
shown in the information box as a light blue line. To select 
a sample, also use the universal knob or the KEYPAD key. 
The pattern flow for all 4 bits will be shown in the
window. Around the index, ±8 bits will be displayed. Use 
the soft menu key VALUE to change the value of the 
selected sample. Use the soft menu key DRAW to draw 
large sample amounts with the same value. The VALUE 
setting remains unchanged and is applied to all selected 
indices. If the pattern is deleted (soft menu key DELETE 
ALL), its length will be reset to 1. All connectors will be 
set to ‚0‘.
 TIMING SETUP:  
Use the soft menu key BIT TIME to set the time at which 
Fig. 10.1: Arbitrary menu
Pin Frequency
S0
f/2
S1
f/4
S2
f/8
S3
f/16
Tab. 10.2: Counter