Lecroy WAVEACE 2022 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2022 用户手册
产品代码
WaveAce 2022
WaveAce 1000/2000
WA1K2K-OM-E RevB
81
Utility Menu - Page 2/4
D
O
S
ELF
C
AL
1. Press the Do Self Cal option button to perform a self-calibration.
The Do Self Cal screen is shown.
2. As the screen indicates, disconnect anything plugged into input
connectors (except for the power cable) before pressing the
SINGLE Front Panel (on page 10) button and performing a self-
calibration. Press the RUN/STOP front panel button to exit the Do
Self Cal screen.
SINGLE Front Panel (on page 10) button and performing a self-
calibration. Press the RUN/STOP front panel button to exit the Do
Self Cal screen.
3. After pressing the SINGLE front panel button, a status bar shows
the self-calibration progress for each channel on the instrument.