29
ACQUISITION - TRIG_SLOPE, TRSL
Description
The TRIG_SLOPE command sets the trigger slope of the specified trigger source. The
TRIG_SLOPE? query returns the trigger slope of the selected source.
TRIG_SLOPE? query returns the trigger slope of the selected source.
Command Syntax
<trig_source>:TRig_SLope <trig_slope>
<trig_source>:= {C1,C2,C3,C4,LINE,EX,EX5}
<trig_slope>:= {NEG, POS}
Query Syntax
<trig_source>:TRig_SLope?
Response Format
<trig_source>:TRig_SLope <trig_slope>
Example
The following instruction sets the trigger slope of Channel 2 to negative:
C2:TRSL NEG
Related Commands
TRIG_COUPLING, TRIG_DELAY, TRIG_LEVEL, TRIG_MODE, TRIG_SELECT,
TRIG_SLOPE
TRIG_SLOPE