Lecroy WAVEACE 1002 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 1002 用户手册

产品代码
WaveAce 1002
下载
页码 123
WaveAce 1000/2000 
 
WA1K2K-OM-E RevB 
81 
Utility Menu - Page 2/4 
 
D
S
ELF 
C
AL
 
1.  Press the Do Self Cal option button to perform a self-calibration. 
The Do Self Cal screen is shown. 
 
2.  As the screen indicates, disconnect anything plugged into input 
connectors (except for the power cable) before pressing the 
SINGLE Front Panel (on page 10) button and performing a self-
calibration. Press the RUN/STOP front panel button to exit the Do 
Self Cal
 screen. 
3.  After pressing the SINGLE front panel button, a status bar shows 
the self-calibration progress for each channel on the instrument.