4.  Analog Channel Threshold Setting 
To judge logic “1” and logic “0” of the buses, you need to set a threshold for 
each analog channel (CH1-CH4). When the signal amplitude is greater than the 
preset value, it is considered as “1”; otherwise “0”.   
Threshold 
Level
1
0
0
 
 
Press MATH  Decode Options  Auto Thre. to turn the auto threshold on 
or off.   
 
When the auto threshold is turned on, the instrument will calculate the center 
point according to the current waveform trace automatically as the threshold of 
each channel before decording. 
 
When the auto threshold is turned off, you can press Thre.Set to turn the auto 
setting menu. Press CH1CH2CH3CH4 and use    to set the threshold of 
each channel. Press 50% to set the current threshold to 50% of the current 
waveform trace. 
 
5.  Display-related Setting 
  Press Format to set the display format of the bus to Hex, Decimal, Binary, 
ASCII or LINE. Note that LINE format is displayed the acyual value of the 
bus in a binary number form, and the order is consistent with the bus 
transmission sequence. This format is only valid for the serial bus, because 
the serial bus has the difference of bit sequence of LSB and MSB; if bit 
sequence of the bus is selected MSB, the LINE format is the same to the 
binary format. 
  Press Pos and use 
  to adjust the vertical display position of the bus. 
 
6.  Decoding Configuration 
Press Decoding1 Configure to turn the decoding configuration submenu on. 
  Press Label to turn the label display function on or off. When it is turned on, 
the bus label will be displayed at the upper left of the bus (when the bus