Rigol DS1104Z-S 4-channel oscilloscope, Digital Storage oscilloscope, DS1104Z-S 用户手册

产品代码
DS1104Z-S
下载
页码 264
RIGOL 
DS1000Z User’s Guide 
1-8 
Function Inspection 
 
1.  Press Storage  Default to restore the instrument to its default configuration. 
2.  Connect the ground alligator clip of the probe to the “Ground Terminal” under 
the probe compensation signal output terminal. 
3.  Use the probe to connect the input terminal of CH1 of the oscilloscope and the 
“Compensation Signal Output Terminal” of the probe.   
 
 
Figure 0-6 To Use the Compensation Signal 
 
4.  Press AUTO. 
5.  Observe the waveform on the display. In normal condition, the display should be 
a square waveform as shown in the figure below: 
 
Figure 0-7 Square Waveform 
 
6.  Use the same method to test the other channels. If the square waveforms 
actually shown do not match that in the figure above, please perform 
 in the next section.   
 
 
WARNING
 
To avoid electric shock during the use of probe, please make sure that 
the insulated wire of the probe is in good condition and do not touch 
the metallic part of the probe when the probe is connected to high 
voltage source. 
Compensation Signal Output Terminal 
Ground Terminal