Example 6: The Application of the X-Y operation 
Viewing Phase Changes through a Network
 
 
Theme: Connect the oscilloscope to monitor the input and output of the circuit and 
capture the phase changes. 
 
To view the input and output of the circuit in an X-Y display, do these steps:
 
1.  From the probe menu set the attenuation to 10X. Set the switch to 10X on the 
probes. 
2.  Connect the CH 1 probe to the input of the network, and connect the CH 2 probe 
to the output. 
3.  If the channels are not displayed, press the CH1 and CH2 buttons. 
4.  Press the AUTO button 
5.  Adjust the vertical
knob to display approximately the same amplitude 
signals on each channel 
6.  Press the MENU in horizontal control area to display the menu 
7.  Press the Time Base soft button to select X-Y 
The oscilloscope displays a Lissajous pattern representing the input and output 
characteristics of the circuit
 
8.  Adjust  the  vertical 
  and 
  knobs  to  a  desirable 
waveform display. 
9.  Apply  the  Ellipse  method  to  observe  the  phase  difference  between  the  two 
channels.   
(See Figure 3-6)