Chauvin Arnoux P01157020 Simple Logger P01157020 用户手册

产品代码
P01157020
下载
页码 172
SIMPLE LOGGER II 
 
 
 
English 
 
 
Page 40 
1.  Inputs (depends on model) 
 
L101/L102: 
1 BNC connector or 2 BNC connectors. 
 
L261/L111: 
2 4 mm safety banana jacks. 
 
L322/L432: 
1 removable 4-position screw-type terminal block. 
 
L562:   
1 BNC connector and 2 4 mm safety banana jacks. 
 
L642:   
2 special miniature thermocouple connectors. 
 
L404:   
One 8-position removable screw-type terminal block. 
 
L702:   
Temp/RH Sensor 
 
ML912:  
Two captive MiniFlex
®
 sensors. 
 
 
2.  Five LED indicators 
The indicators have two distinct information functions: choice of command to be executed or active function 
in progress. 
The command to be executed (reached by pressing the "PRESS" button until the desired command appears) 
is indicated to the left of the indicator. 
The active function or current status (obtained by a brief press on the "PRESS" button) is indicated to the 
right of the indicator. 
Section 4.1 describes each of the indicators in detail. 
 
3.  Command button (« PRESS ») 
The button marked "PRESS" is used to choose the operating mode: start or stop recording, erase the 
memory, clear alarms, start or stop the device. 
 
4.  Mini B female USB connector 
This connector, on the base of the device, is used for connection to a desktop computer. 
 
5.  « RESET »  button(not shown) 
The "RESET" button is used to reset the processor. This button is reached by opening the housing: remove 
the battery compartment cover and unscrew the four housing closing screws; the button is located on the 
LED indicators side of the circuit, near the inputs. 
 
6.  "Flash Upgrade" switches (not shown) 
These two switches, placed near the "RESET" button, are used to take control of the device if an upgrade of 
the internal operating program fails. 
 
 
 
WARNING:  if the "RESET" button is actuated while recording is in progress, recording will be 
interrupted and the data in memory may be lost.