Fluke 1653B05-TPOLEKIT/FVDE-tester 4426033 用户手册

产品代码
4426033
下载
页码 66
 Electrical 
Installation 
Tester 
 Specifications 
51 
Insulation Resistance Measurement (R
ISO
Test Voltages 
Model 1652C 
Model 1653B 
Model 1654B 
Accuracy of Test 
Voltage 
(at rated test 
current) 
250-500-1000 V 
50-100-250-500-1000 V 
+10 %, -0 % 
 
Test 
Voltage 
Insulation 
Resistance Range 
Resolution
Test Current 
Accuracy 
50 V 
10 k
Ω
 to 50 M
Ω
 0.01 
M
Ω
 
1 mA @ 50 k
Ω
 
±
(3 % + 3 digits) 
100 k
Ω
 to 20 M
Ω
  
0.01 M
Ω
 
±
(3 % + 3 digits) 
100 V 
20 M
Ω
 to 100 M
Ω
 0.1 
M
Ω
 
1 mA @ 100 k
Ω
 
±
(3 % + 3 digits) 
10 k
Ω
 to 20 M
Ω
 0.01 
M
Ω
 
±
(1.5 % + 3 digits) 
250 V 
20 M
Ω
 to 200 M
Ω
 0.1 
M
Ω
 
1 mA @ 250 k
Ω
 
±
(1.5 % + 3 digits) 
10 k
Ω
 to 20 M
Ω
 0.01 
M
Ω
 
±
(1.5 % + 3 digits) 
20 M
Ω
 to 200 M
Ω
 0.1 
M
Ω
 
±
(1.5 % + 3 digits) 
500 V 
200 M
Ω
 to 500 M
Ω
 1 
M
Ω
 
1 mA @ 500 k
Ω
 
±
10 % 
100 k
Ω
 to 200 M
Ω
 0.1 
M
Ω
 
±
(1.5 % + 3 digits) 
1000 V 
200 M
Ω
 to 1000 M
Ω
 1 
M
Ω
 
1 mA @ 1 M
Ω
 
±
10 % 
Note 
The number of possible insulation tests with a fresh set of batteries is 2000. 
 
Auto Discharge 
Discharge time constant <0.5 second for C = 1 
μ
or less. 
Live Circuit Detection 
Inhibits test if terminal voltage >30 V prior to 
initiation of test. 
Maximum Capacitive Load 
Operable with up the 5 
μ
F load.