6.  HOLDTHRESH 
Command Format: 
HOLDTHRESH <threshold> 
HOLDTHRESH? 
 
Function: 
The commands set and query the threshold for HOLD measurement. 
 
Explanations: 
HOLDTHRESH  <threshold>  -  Sets the HOLD threshold to <threshold>, 
<threshold> should be 1, 2,  3 or 4 (which  are 0.01%, 0.1%, 1%, 10%, 
respectively), or else errors may occur. 
HOLDTHRESH? - Queries the threshold of HOLD measurement. 
 
Return Format: 
The query returns the HOLD threshold 1, 2, 3 or 4. 
 
 
7.  MAX 
Command Format: 
MAX 
 
Function: 
The command enables the multimeter enter MAX operation with current reading 
as maximum. 
 
Explanations: 
If the multimeter has been in MAX mode, the maximum value will be shown on 
the screen. Note the auto range function is disabled in this mode.