Rigol DM3058 Digital-Multimeter, DMM, DM3058 数据表

产品代码
DM3058
下载
页码 192
RIGOL 
Programming Guide for DM3058/DM3058E 
 
                                        
6-3 
The programming implementations——Agilent Commands 
0    cmdset agilent             /* Select Agilent commands */ 
1    *CLS   
 
 
 
 
/* Clear the instrument state */ 
2    *IDN?   
/* Query the equipment ID and confirm if the multimeter works normally */ 
3  /*Return Format: 
RIGOL Technologies,DM3058,DM3A020080808,99.00.00.00.00.00*/ 
4    conf:volt:dc auto/*Enable the DCV measurement and select auto range */ 
5    calc:stat on  
 
        /* Enable the math function */ 
6    calc:func aver   
        /* Select statistic function */ 
7    calc:aver:max?   
  /* Query the current maximum of DCV measurement */ 
8    /*Return Format:5.000104e-02*/ 
9    calc:aver:coun?   
      /* Query the current times of Auto measurement */ 
10   /*Return Format:63*/ 
11   calc:stat OFF             /* Exit the math function */ 
 
Explanations: 
1.  The default measurement rate changes into “Fast” after the command in line 4 
was executed, as it is directly corresponding to the resolution, which means you 
can get a desired rate by  setting the precision. For example, sending 
“conf:volt:dc:auto,max” to change the rate to “slow”. 
2.  The command in line 7 returns the current measured maximum of DC voltage. 
3.  The command in line 9 returns the measured times that have been completed.