Programming Guide for DM3058/DM3058E 
6-12 
The programming implementations——Agilent Commands 
0    *CLS   
 
 
 
 
/* Clear the instrument state */ 
1     cmdset agilent             /**Select Agilent commands */ 
2    *IDN?   
 
 
 
 
/*  Query the equipment ID and confirm if  the 
multimeter works normally */ 
3     /* Connect the test leads for DC voltage measurement */ 
4    conf:volt:dc <range> 
    /* Set the DC voltage measuring function and the 
range */ 
5    calc:stat on               /* Enable the math function */ 
6    calc:lim:low <value1>      /* Set the lower limit to value1*/ 
7    calc:lim:upp <value2>      /* Set the upper limit to value2*/ 
8    calc:func lim               /* Enable the limit operation */ 
9    calc:stat OFF               /* Exit math function */ 
 
Explanations: 
1.  The instrument should be connected under DC voltage mode before sending the 
command in line 4. 
2.  Parameters in line 6 and 7 should be: value2 > value1, and both are within the 
range. 
3.  Command in line 8 enables limit operation without return value, as  no 
corresponding command is provided in this system. If the parameter input is not 
within the scope of the upper and lower limits, a beep tone will be emitted ( when 
the beeper is on). 
4.  Use RIGOL commands if both the P/F measurement function and returned value 
are required.