PCS-1000 User Manual 
38
 
Setting the Averaging Mode 
 
Background 
There are two different types of averaging 
modes, SHIFT or TOTAL. 
SHIFT is a box car averaging mode while 
TOTAL will average all the collected samples 
to get the average value.  
 
  Range 
SHIFT, TOTAL 
Steps 
1.  Long push the Func key. 
The function menu will appear.  
 
2.  Use the ◄ Func ► keys to navigate to the AVG 
MODE menu. 
 
3.  Use the ▼Select ▲ keys to select the Averaging 
Mode. Press the Enter key to set the mode. 
By default the average mode is set to SHIFT. 
 
4.  Use the ◄ Func ► keys to navigate to the 
SAVE FUNC SET menu item. Press the Enter 
key to save.  
 Note 
To exit without saving, navigate to the 
EXIT FUNC 
SET
 menu using the ◄ 
Func
 ► keys and press the 
Enter
 key to exit without saving any settings.