Quick Guide for DM3058/DM3058E 
III 
E
N
G
L
IS
H
 
2)  LO-ground  protection  limitation.  LO  terminal can safely 
“float” 500Vpk relative to the ground. 
The maximum protection limitation of HI terminal relative to 
the  ground is 1000Vpk.  Therefore, the sum of the “float” 
voltage and the measured voltage cannot exceed 1000Vpk. 
2  Sampling (HI Sense and LO Sense) terminal 
HI Sense  and  LO Sense  are used for 4-Wire Resistance 
Measurement. Two protection limitations are defined: 
1)  HI Sense-LO Sense protection limitation: 200Vpk. 
2)  LO Sense-LO protection limitation: 2Vpk. 
3  Current input (I) terminal 
I  and  LO  terminal  are used for current measurement. The 
maximum current which go through the I terminal is limited to 
10A by the fuse on the rear panel. 
NOTE: Voltage on the current input terminal corresponds to voltage 
on  LO  terminal.  To obtain favorable  protection, specified fuse 
should be used.   
IEC Measurement Category II Overvoltage Protection 
To protect against the danger of electric shock, DM3058/DM3058E 
provides overvoltage protection for line-voltage mains connections 
meeting both of the following conditions: 
1  The HI and LO input terminals are connected to the mains 
under Measurement Category II conditions, defined below. 
2  The mains are limited to a maximum line voltage of 600VAC. 
WARNING:  IEC Measurement Category II  includes electrical 
devices connected to mains at an outlet on a branch circuit. Such 
devices include most small appliances, test equipment, and other 
devices that plug into a branch outlet or socket.   
DM3058/DM3058E may be used to make measurements with the HI 
and LO inputs connected to mains in such devices (up to 600VAC), 
or to the branch outlet itself. However, DM3058/DM3058E may not 
be used with its HI and LO inputs connected to mains in 
permanently installed electrical devices such as the main 
circuit-breaker panel, sub-panel disconnected  boxes, or 
permanently wired motors. Such devices and circuits are subject to 
overvoltage that may exceed the protection limits of 
DM3058/DM3058E. 
NOTE:Voltages above 600VAC may be measured only in circuits 
that are isolated from mains. However, transient overvoltage is also 
present on circuits that are isolated from mains. DM3058/DM3058E 
is designed to safely withstand occasional transient overvoltage up 
to 4000Vpk. Do not use this equipment to measure circuits where 
transient overvoltage could exceed this level.