User’s Guide for DM3068 
 
2-79 
To Set the Configuration 
 
Press 
   System  Cfg to enter the following interface.     
 
 
1.  PwrOn 
Select a system configuration to be used at power-on from “Default” and 
“Last (configuration at last power-off). 
  This setting will be available at the next power-on.   
  The multimeter will always use DCV as the default measurement at every 
power-on no matter Default or Last is selected. 
 
2.  Default 
Restore the multimeter to factory defaults (see Table 2-8).   
 
3.  Switch 
Enable or disable the front power button. The default status is OFF. 
  ON: enable the front power button. After turning on the power switch at 
the rear panel, users need to press the power button at the front panel to 
turn on the multimeter.   
  OFF: disable the front power button. Users can directly start the 
multimeter by turning on the power switch at the rear panel. However, 
you can still turn off the multimeter by pressing the power button at the 
front panel. 
 
Table 2-8 Factory defaults 
Parameter 
Factory Default   
Measurement Parameters 
*Short-circuit Resistance 
10 Ω 
AC Filter 
Mid 
DC Input Impedance 
10 MΩ 
Reading Resolution 
6 ½ 
Measurement Function 
DCV