Rigol DM3068 Digital-Multimeter, DMM, DM3068 数据表

产品代码
DM3068
下载
页码 171
RIGOL 
User’s Guide for DM3068 
6-8 
10.00000 A
[6]
 
3 Hz-5 Hz 
1.10 + 0.08   
1.10 + 0.10    1.10 + 0.10   
0.100 + 0.008   
5 Hz-10 Hz 
0.35 + 0.08   
0.35 + 0.10    0.35 + 0.10   
0.035 + 0.008   
10 Hz-5 kHz 
0.15 + 0.08   
0.15 + 0.10    0.15 + 0.10   
0.015 + 0.008   
 
Additional Low Frequency Errors (% of reading) 
Additional Crest Factor Errors (non-sine wave)
 [7]
 
Frequency 
AC Filter 
Crest Factor 
Error (% of reading) 
Slow 
Medium 
Fast 
10 Hz-20 Hz 
0.74 
-- 
1 - 2 
0.05 
20 Hz-40 Hz 
0.22 
-- 
2 - 3 
0.2 
40 Hz-100 Hz 
0.06 
0.73 
3 - 4 
0.4 
100 Hz- 200 Hz 
0.01 
0.22 
4 - 5 
0.5 
200 Hz-1 kHz 
0.18 
 
>1 kHz 
[1] 
Specifications are for 90-minute warm-up, slow ac filter and sinewave input. 
[2] 
10% overrange on all ranges except ACV 750 V and ACI 10 A ranges. 
[3] 
Relative to calibration standards. 
[4] 
Specifications are for sinewave input >5% of range. For inputs from 1% to 5% of range and <50 kHz, add 0.1% of range additional error. For 50 kHz 
to 100 kHz, add 0.13% of range. 
[5] 
ACV 750 range limited to 8x10
Volt-Hz. When the input exceeds 300 V rms, add 0.7 mV error for each additional volt. 
[6] 
For continuous current > DC 7A or AC 7A rms, 30 seconds ON and 30 seconds OFF. 
[7] 
When the frequency is lower than 100 Hz, the specification of slow filter is only for sine wave input. 
[8] 
Specifications are for sinewave input >5% of range. For inputs from 1% to 5% of range, add 0.1% of range additional error. Specifications are typical 
values for 200 uA, 2 mA, 2 A and 10 A ranges when frequency is >1 kHz.